4335 - Vol. 78 / Ed 4 / in 2012
Section: Artigo Original Pages: 120 to 125
From nystagmus to the air and water caloric tests
Authors:
Anna Carolina Marques Perrella de Barros1; Heloisa Helena Caovilla2

Keywords: caloric tests, labyrinthine fluids, nystagmus, physiologic, vestibular function tests, vestibule, labyrinth.

Abstract: The caloric test is an important tool for the assessment of labyrinthine function.
OBJECTIVE: To compare the nystagmus response in the caloric tests with air at 50ºC and 24ºC and with water at 44ºC and 30ºC. Study Design: Randomized crossover clinical trial.
MATERIALS AND METHODS: 40 healthy individuals were submitted to a neurotological evaluation, including caloric tests with air at 50ºC and 24ºC and water at 44ºC and 30ºC.
RESULTS: Comparing the air and water caloric tests, there were no significant differences among the post-caloric nystagmus slow-phase velocity in relation to the stimulation order, between ears and between the values of unilateral weakness and directional preponderance. The slow-phase velocity values were higher with water (p = 0.008, p < 0.001), and cold stimulation produced stronger responses (p < 0.001).
CONCLUSION: Comparing 50ºC and 24ºC air caloric test and 44ºC and 30ºC water caloric test, we observed similar slow-phase velocity values for both ears, higher responses in the cold temperature and in the test with water, and similar results of unilateral weakness or directional preponderance for post-caloric nystagmus in both tests.

INTRODUCTION

The caloric test is usually the most informative part of vestibular evaluation. It is the only procedure which assesses each labyrinth separately, it informs about the affected side and enables characterization of the vestibular lesion intensity1,2, despite having large intra and inter-individual variability in its evaluation parameters2,3.

Heat stimulation of the labyrinth causes endolymphatic current in the semicircular canals, polarizing or depolarizing the sensorial cells of the ampule crest, triggering the vestibular-ocular reflex. The stimulus generates an ampule current, towards the utricle, exciting the stimulated lateral semicircular canal and causing nystagmus in the same direction of the labyrinth being tested. The cold stimulus results in a current in the opposite direction of the utricle, inhibiting the stimulated semicircular canal and nystagmus in the opposite direction of the assessed labyrinth2-5.

Besides the convection currents, other mechanisms may be involved in evoking caloric responses, such as the mechanical transduction of the cupule, the heat effect on sensorineural structures and neural adaptation, combined in a non-linear fashion2.

Under physiological conditions, the vestibular organs are sensitive to stimuli varying between 0.001 Hz and 8 Hz6, and the semicircular canals respond more efficiently to angular movements between 1 and 6Hz7.

The caloric test enables the assessment of the lateral semicircular canal; defects must be interpreted together with other tests, since they do not necessarily indicate that the labyrinth is involved8. This tests provide partial information about the vestibular function, because it causes an endolymphatic current in the corresponding lateral semicircular canal at only 0.003 Hz of the head angular movement7.

Many methods have been created in order to assess vestibular function by means of heat stimulation. Initially, water was the only element employed to evoke the nystagmic response9. Ether and ethyl chloride10 have also been used, and the biphasic caloric test with air has been advocated, in which hot and cold temperatures are quickly alternated during the stimulation, producing a bidirectional nystagmus11; the caloric test with continuous thermal change, in which the temperature is regulated during the test and reduced at a constant rate until nystagmus ensues12; and one with a silicone balloon in the external acoustic meatus with circulating water, which enables the test to be carried out in ears with perforated ear drums13; stimulation could be alternate bithermal, simultaneously bithermal, monothermal and by means of cold tests only2.

Air irrigators became commercially available around the 60's14 and are broadly used today. In Brazil, one air caloric stimulation test equipment used, the Ototermoar, caused a reaction in the labyrinth similar to that with water in healthy individuals and differentiated healthy individuals from patients with vestibular disorders, enabling the conclusion that the air test was technically doable, and could be used in clinical practice15.

Air is more comfortable and safer for the test, without contraindications to its use, and it can be used in the assessment of patients with eardrum perforations, patients with external otitis media and mastoidectomy cavities2,10,16-18. Differently from water, it does not need to be collected during irrigation and the necessary time to reach the warmer temperature is shorter9.

The thermal properties of air and water are significantly different. The capacity to maintain the air temperature is lower than that of water19,20. The technical demand for the test with air is higher16, though. The model and diameter of the irrigator tip, how much is inserted into the external acoustic meatus, the direction and flow of air, the room temperature and the location of the equipment in the testing room may affect the characteristics of the air; thus, the amplitude of post-caloric responses3,16,17,21.

The literature discusses the effort of researchers to define the best procedure and to obtain the caloric response with air. The temperatures utilized in the caloric test with hot air varies between 42ºC and 51ºC and in cold test, they vary between 20ºC and 30ºC, with air flows between 5 and 13L/min and stimulation time between 60 and 100 seconds10,14,17,18,20,22,23.

In the irrigation with water, the temperatures employed are 7ºC higher and lower than body temperature; 44ºC for the warm test and 30ºC for the cold one. In the irrigation with air, one needs gradients of 13ºC, corresponding to 50ºC, for the hot stimulation and to 24ºC for the cold one, in order to evoke a response with amplitude similar to the one obtained by water stimulation2,16,23.

The most important parameter for the caloric test quantitative assessment, which can be analyzed thanks to the possibility of recording the response, is the maximum Slow Component Angular Velocity (SCAV), which is proportional to the intensity of the employed stimulus24. The hot and cold stimulations with air usually produce nystagmic responses which are similar to the ones produced with water19,25, with response reliability similar to that of the test with water9,19 and proper test-retest reliability26. Nonetheless, there is a report that the absolute SCAV values were more intense in the test with water17,20 and showed greater variability in the test with air22,27. Different responses in the comparison between hot and cold stimuli are not very important, since the prime goal of the caloric test is to show whether the response to the same hot or cold stimulus is symmetrical in both ears28.

The caloric stimulation with air was challenged because of response variability9,16,27. The variability of post-caloric responses with air could be explained by the tendency of air temperature balancing out with the room temperature. The use of temperatures farther from the body temperature and the increase in irrigation time and flow may broaden response intensity; the test-retest variations drop when the stimulus intensity and flow are increased, and when the temperature is closer to that of the room22. At levels above 10 L/min, the discomfort caused by the acoustic stimulus is higher22 and, at levels below 5 L/min, the heat transfer is markedly reduced23.

Recent studied on the caloric test with water and air usually approach topics associated with the findings in vestibular disorders. In our country, in caloric tests with air at 42ºC and 18ºC and with water at 30ºC and 44ºC in chronic peripheral vestibular disorders, they found a prevalence of cases with SCAV values abnormalities on labyrinthine predominance and directional preponderance29.

The diagnostic importance of the caloric test and the scarce information in our literature on the comparison between thermal labyrinthine stimulations with water and with air motivated this study.

The temperatures of 50ºC and 24ºC of the caloric test with air has been recommended by advanced neurotology centers of numerous countries; and in our country there is no scientific paper comparing the results of this test with the ones from the conventional caloric test with water.

The present study can be very practical for the routine assessment of vestibular function and it also opens horizons for other studies with the caloric test with air in these temperatures, enabling a pertaining comparison with the results from the international literature.

The goal of the present study is to compare the post-caloric nystagmus of the test with air at 50ºC and 24ºC with that from the test with water at 44ºC and 30ºC.

MATERIALS AND METHODS

The present study was carried out after approval by the Ethics in Research Committee of the institution, under protocol number 0054/09. The informed consent form was signed by all the subjects before the study onset.

The study group was made up of 40 volunteers, 25 females and 15 males, with ages between 18 and 40 years, without ear complaints.

As for inclusion criteria, we selected healthy individuals without signs and symptoms of body balance disorder and who were not using medication. We carried out a guided interview to rule out a past of neurotological diseases and use of medication. We also excluded healthy individuals with changes in the vestibular test with spontaneous nystagmus with eyes closed and/or pre-caloric, because of the possible influence of these signs on the caloric test.

After visually examining the external ear canal, to make sure there was no wax or foreign body, the subjects were submitted to vestibular function assessment4, made up of static and dynamic balance tests; positional and positioning nystagmus study; and study of spontaneous and semi-spontaneous nystagmus, saccadic movements, pendular tracking, optokinetic nystagmus, decreasing pendular rotational test and caloric test with air at 50ºC and 24ºC and with water at 44ºC and 30ºC with digital vector electronystagmography (VENG - Neurograff Eletromedicina Ind. e Com. Ltda. - EPT - Brazil), using the otocalorimeter with air (model OAT 10, adapted for temperatures of 50ºC and 24ºC, Neurograff Eletromedicina Ind. e Com. Ltda. - EPT - Brazil) and otocalorimeter with water (model OC 214 - Berger). The responses were registered and analyzed in the VecWin 2 (Neurograff Eletromedicina Ind. e Com. Ltda - EPT - Brazil) software. Before data collection, the otocalorimeters were calibrated, with temperature adjustments according to the technical specs from the manufacturers.

The individuals were distributed in two groups in a randomized fashion, according to the caloric test, which was carried out first. Group I was initially submitted to the caloric test with air and Group II was initially submitted to caloric test with water, in order to avoid a possible influence in the order of stimulations, habituation and residual effects of one test on the other23.

Before the functional assessment of the vestibular system, the patients were instructed to refrain from smoking, eating chocolate, drinking alcoholic or caffeine loaded beverages in the three days before the assessment. On the day of the test, they were instructed to refrain from using makeup, facial cremes or contact lenses.

In the exam room, the lighting was dimmed, and the room had air conditioning, keeping the room temperature at about 21ºC, to avoid interference from air and water thermal variations in the results.

In order to carry out the VENG, the patients had their skins cleaned in order to enable the placement of three active silver chloride leads and one ground lead. The leads were placed in the right external periorbital corner, in the left external periorbital corner and in the frontal midline, using the triangular placement of the leads which enables the identification of nystagmus direction and to check the slow component angular velocity (SCAV) in three recording channels30.

The calibration of eye movements was done in the beginning and, when necessary, throughout the test and, systematically before the caloric test to minimize changes to the response amplitude caused by the cornea-retina electrical variation.

During the caloric test with water and air, the subject was kept in the supine position with the head tilted at 30ºC31. The heat stimulation was carried out in each ear separately, with air at 50ºC and 24ºC, in a 8L/min flow, for 60 seconds9,19, keeping a three-minute interval between stimulations19; and with 250 ml of water at 44ºC and 30ºC, during 40 seconds, with a five-minute interval between stimulations4,32. We had a 15-minute interval after the end of the nystagmus response from the last stimulation with caloric test with water or with air and the onset of the first stimulation of the next test with air or water.

Each subject was submitted to eight hot and cold caloric stimulations with air or water, carried out in the following order: hot in the right ear, hot in the left ear, cold in the left ear and cold in the right ear. We used strategies such as conversation, naming and the use of mathematical calculations to stimulate mental activity and to avoid the cortical inhibition of the caloric response. Vertigo, SCAV and direction of the post-caloric nystagmus were analyzed with the eyes closed and eyes open. SCAV was chosen as the parameter to assess response intensity.

The entire post-caloric nystagmus recording was assessed by selecting the most representative stretch of the post-caloric nystagmus record, with the most intense responses in terms of SCAV. The percentage values of the labyrinthine predominance and directional preponderance of the post-caloric nystagmus were calculated using the SCAV values in the four stimulations with air and in the four stimulations with water33.

We did a statistical descriptive analysis to characterize the sample. For the quantitative variables we observed minimum and maximum values and calculated the mean and standard deviation values. For the qualitative variables we calculated the absolute and relative frequencies.

We used the chi-square test to assess the homogeneity between the groups I (air-water) and II (water-air) vis-à-vis the age and gender variables.

The t-Student test was utilized in the comparative analysis of Groups I (air-water) and II (water-air) vis-à-vis the mean SCAV value of the post-caloric nystagmus.

We used the paired t-Student test in the comparative analysis between the right and left ears in the tests with air at 50 and 24ºC and with water at 44ºC and 30ºC, in the comparative analysis by ear between the tests with air, at 50ºC and 24ºC, and with water at 44ºC and 30ºC, in the comparative analysis between the tests with air at 50ºC and 24ºC and with water at 44ºC and 30ºC in each ear and in the comparative analysis between the relative values of labyrinthine predominance and nystagmus directional preponderance in the tests with air at 50ºC and 24ºC and with water at 44ºC and 30ºC.

The statistical analyses were done by the SPSS 15.0 for Windows (Statistical Package for Social Sciences, version 15.0, 2006) software; the significance level employed was 5% (α = 0.05). The power of the tests was calculated when there was significant difference between the groups; the values found varied between 87% and 100.0%, indicating that the sample size was enough to obtain more than 80% of power.

RESULTS

We assessed 40 healthy individuals, 25 (62.5%) females and 15 (37.5%) males, with mean age of 26.45 years (ranging between 18 and 39 years) and standard deviation of 5.62 years. All the individuals reported having vertigo upon caloric stimulation with air and with water.

Table 1 shows the descriptive values of the distribution of the healthy individuals in Groups I (air-water) and II (water-air) and the comparative analysis between the groups as to age and gender. There was no statistically significant difference between the two groups vis-à-vis age and gender.




Table 2 depicts the mean, standard deviation, minimum and maximum values of the slow component angular velocity of the post-caloric nystagmus in the tests with air at 50ºC and 24ºC and with water at 44ºC and 30ºC in the right and left ears, and the p-value of the comparative statistical analysis, considering the order at which the tests were performed. Regardless of such order, there were no statistically significant differences between the post-caloric nystagmus SCAV values with air and there were no statistically significant differences between the post-caloric nystagmus SCAV with water.




Table 3 depicts the mean, standard deviation, minimum and maximum SCAV values and labyrinthine predominance and directional preponderance of post-caloric nystagmus upon stimulation with air at 50ºC and 24ºC and with water at 44ºC and 30ºC, and the comparative statistical analysis of the results between the ears, the temperatures and the tests. There was no statistically significant difference upon comparison of the SCAV values between the ears in the caloric tests with air and water. There was a statistically significant difference (p < 0.001) between the SCAV values upon comparison of the test with hot air with that of cold air and upon comparing the hot water with the cold water tests in each ear; the cold temperature evoked more intense responses than the hot temperature in the tests with air and water. There was a statistically significant difference (p = 0.008; p < 0.001) between the caloric test with air in relation to the test with water in all the stimulations; the SCAV values were higher in the stimulations with water than air. There were no statistically significant differences between the values of labyrinthine predominance (p = 0.761) and directional preponderance (p = 0.391) upon comparison of the results from the tests with water and air.




DISCUSSION

Comparing the results from the air-water (Group I) and water-air (Group II) groups, as to the order in which the tests were carried out, there was no statistically significant difference between the post-caloric nystagmus SCAV with air and between the post-caloric SCAV values with water, regardless of which test preceded the other, ruling out a possible effect of the order of post-caloric responses. Similarly to our findings, in another study, there was no evidence of post-caloric nystagmus physiological adaptation in normal individuals, because the order in which the stimulations were done also did not impact response amplitude34.

There were no statistically significant differences when we compared the responses from the right ear with those from the left ear in the stimulations at 50ºC and 24ºC in the caloric test with air and at 44ºC and 30ºC with water, similarly to what has been reported by other authors17,23.

We noticed that SCAV values were higher with water stimulation when compared to those with air, for both ears. Other authors have also reported higher evoked post-caloric responses with water than air14,20,26 or similar responses10,17,19,23.

There was a statistically significant difference in the comparative analysis of the hot stimulations (50ºC with air and 44ºC with water) and the stimulations with cold temperatures (24ºC with air and 30ºC with water) in each ear. The cold temperature caused more intense responses than the hot one in both tests: with air and with water. Similarly to our findings, there are reports of higher responses in the cold temperature than in the hot one using water26,28,35 and air26. Notwithstanding, other studies have shown that the hot stimulus caused more intense responses than its cold counterpart in the caloric test with water9,14 and with air9,11, or that similar results were obtained with both temperatures in the tests with air14,17,19,23 and with water17,19,23.

There was no significant difference between the relative values of labyrinthine predominance and nystagmus directional preponderance in the caloric tests with air and water, similarly to what was found in other studies14,23. The classical indicator of peripheral vestibular lesions is the comparison of the responses from both ears in the bithermal caloric test, by using the labyrinthine predominance formula22. Using air as a caloric stimulus does not impact the clinical measure of labyrinthine predominance27.

The use of different configurations of flow, duration and temperature of the thermal stimulus in the assessment of the vestibular function10,14,17,18,20,22,23, and the use of different protocols to perform the caloric test with water and air, make it difficult to compare and to integrate the findings from many research centers, thus being necessary that each laboratory set its own reference values36.

The differences found in the post-caloric responses upon comparison with other studies in healthy individuals may be justified by the different thermal properties of water and air as vectors of labyrinthine stimulation. Different responses are expected when using water, since it has a higher specific heat than air and it is capable of keeping the temperature in the external ear canal for longer; air satisfactorily replaced water in 95% of the clinical cases seen in a hospital; the remaining 5% had lower labyrinthine function, confirmed with the use of a more intense stimulus: ice water19.

Considering the equivalence between the relative values of labyrinthine predominance and post-caloric nystagmus directional preponderance in the test with air and water in the healthy individuals we assessed, and that such parameters are fundamental to identify vestibular dysfunctions, we can state that both tests are useful in neurotological routine testing.


CONCLUSION

Comparing the tests with air at 50ºC and 24ºC and with water at 44ºC and 30ºC, we found two similar values for the slow component angular velocity in both ears to be higher with cold temperature and in the test with water, and similar results of labyrinthine predominance and post-caloric nystagmus directional preponderance in both tests.


REFERENCES

1. Baloh RW, Furman JM. Modern vestibular function testing. West J Med. 1989;150(1):59-67.

2. Ganança MM, Bottino MA, Bittar RSM, Caovilla HH, Ganança FF. Reference standard to read the air-driven caloric reflex test results. Braz J Otorhinolaryngol. 2009;75(1):2.

3. Jacobson GP, Newman CW. Background and technique of caloric testing. In: Jacobson GP, Newman CW, Kartush JM. Handbook of balance function testing. St. Louis: Mosby; 1993. p.156-90.

4. Ganança MM, Caovilla HH, Munhoz MSL, Silva MLG, Frazza MM. As etapas da equilibriometria. In: Caovilla HH, Ganança MM, Munhoz MSL, Silva MLG. Equilibriometria clínica. São Paulo: Atheneu; 1999. p.41-97.

5. Honrubia V. Testes quantitativos da função vestibular e o exame clínico. In: Herdman SJ. (Org). Reabilitação vestibular. 2ª ed. São Paulo: Manole; 2002. p.105-68.

6. Bittar RSM, Bottino MA, Pedalini MEB, Ramalho JRO, Carneiro CG. Arreflexia pós-calórica bilateral: aplicabilidade clínica da reabilitação vestibular. Rev Bras Otorrinolaringol. 2004;70(2):188-93.

7. Gonçalves DU, Felipe L, Lima TMA. Interpretation and use of caloric test. Braz J Otorhinolaryngol. 2008;74(3):440-6.

8. Wuyts FL, Furman J, Vanspauwen R, Van de Heyning P. Vestibular function testing. Curr Opin Neurol. 2007;20(1):19-24.

9. Ford CR, Stockwell CW. Reliabilities of air and water caloric responses. Arch Otolaryngol. 1978;104(7):380-2.

10. Albernaz PL, Ganança MM. The use of air in vestibular caloric stimulation. Laryngoscope.1972;82(12):2198-203.

11. Fleming PM, Proctor LR, Dix RC, Metz WA. Results of new air caloric testing among normal subjects. I. Biphasic testing. Ann Otol Rhinol Laryngol. 1978;87(2 Pt 1):248-56.

12. Itaya T, Kitahara M. Air caloric test with continuous thermal change in patients with vestibular disorders. Acta Otolaryngol. 1995;519:184-7.

13. Henry DF. Test-retest reliability of open-loop bithermal caloric irrigation responses from healthy young adults. Am J Otol. 1999;20(2):220-2.

14. Zapala DA, Olsholt KF, Lundy LB. A comparison of water and air caloric responses and their ability to distinguish between patients with normal and impaired ears. Ear Hear. 2008;29(4):585-600.

15. Mangabeira Albernaz PL, Ganança MM, Imperatriz A, Lamoglia Neto J. Ototermoar. Ars Curandi. 1971;12:58-78.

16. Proctor LR. Air caloric test: irrigation technique. Laryngoscope. 1977;87(8):1383-90.

17. Benitez JT, Bouchard KR, Choe YK. Air calorics: a technique and results. Ann Otol Rhinol Laryngol. 1978;87(2 Pt 1):216-23.

18. Gao YZ, Sze YY, Shen L. The air caloric test and its normal values. Adv Otorhinolaryngol. 1983;31:191-7.

19. Capps MJ, Preciado MC, Paparella MM, Hoppe WE. Evaluation of the air caloric test as a routine examination procedure. Laryngoscope. 1973,83(7):1013-21.

20. Greven AJ, Oosterveld WJ, Rademakers WJ, Voorhoeve R. Caloric vestibular test with the use of air. Ann Otol Rhinol Laryngol. 1979;88(1 Pt 1):31-5.

21. Lightfoot G, Milner L. The dependency of air caloric stimulus effectiveness on delivery tip characteristics. Int J Audiol. 2010;49(10):772-4.

22. Coats AC, Hebert F, Atwood GR. The air caloric test. A parametric study. Arch Otolaryngol. 1976;102(6):343-54.

23. Tole JR. A protocol for the air caloric test and a comparison with a standard water caloric test. Arch Otolaryngol. 1979;105(6):314-9.

24. Henriksson NG. Speed of slow component and duration in caloric nystagmus. Acta Otolaryngol Suppl. 1956;125:1-29.

25. Barber HO, Stockwell CW. Manual of Electronystagmography. 2nd ed. St. Louis: Mosby; 1980. p.230.

26. Karlsen EA, Mikhail HH, Norris CW, Hassanein RS. Comparison of responses to air, water, and closed-loop caloric irrigators. J Speech Hear Res. 1992;35(1):186-91.

27. Cooper JC Jr, Mason RL. Variability of air calorics vs water. Statistical implications. Arch Otolaryngol. 1979;105(3):113-5.

28. Maire R, Daoui B, Van Melle G. Evaluation of the caloric test by combining 3 response parameters. Otolaryngol Head Neck Surg. 2000;122(6):814-20.

29. Bovolini A, Ganança CF, Ganança FF, Ganança MM, Caovilla HH. Prevalência de anormalidades às provas calóricas com água e com ar em vestibulopatias periféricas crônicas. Acta ORL. 2007;25(2):165-9.

30. Ganança MM, Caovilla HH, Ganança FF. Electronystagmography versus videonystagmography. Braz J Otorhinolaryngol. 2010;76(3):399-403.

31. Hallpike CS. The caloric tests. J Laryngol Otol. 1956;70(1):15-28.

32. Cruz NA, Ganança MM. Avaliação da prova térmica de Fitzgerald & Hallpike em 101 indivíduos normais com e sem electronistagmografia. Hospital (Rio J). 1968;73(2):613-41.

33. Jongkees LB, Philipszoon AJ. Electronystagmography. Acta Otolaryngol Suppl. 1964;189:Suppl 189:1+.

34. Lightfoot GR. The origin of order effects in the results of the bi-thermal caloric test. Int J Audiol. 2004;43(5):276-82.

35. Karlsen EA, Hassanein RM, Goetzinger CP. The effects of age, sex, hearing loss and water temperature on caloric nystagmus. Laryngoscope. 1981;91(4):620-27.

36. Maes L, Dhooge I, De Vel E, D'haenens W, Bockstael A, Vinck BM. Water irrigation versus air insufflation: a comparison of two caloric test protocols. Int J Audiol. 2007;46(5):263-9.










1. MSc in Sciences - Graduate Program in Human Communication Disorders - Federal University of São Paulo - Paulista School of Medicine (UNIFESP/EPM) (Speech and Hearing Therapist).
2. Senior Associate Professor of Otology and Neurotology - Federal University of São Paulo - Paulista School of Medicine (UNIFESP/EPM) (Speech and Hearing Therapist).

Federal University of São Paulo - Paulista School of Medicine (UNIFESP/EPM).

Send correspondence to:
Anna Carolina Marques Perrella de Barros
Rua Baguassu, 141, apto 173. Vl
Regente Feijó - SP. CEP: 03344-015.
Email: anna_perrella@yahoo.com.br

Paper submitted to the BJORL-SGP (Publishing Management System - Brazilian Journal of Otorhinolaryngology) on January 15, 2012.
Accepted on March 03, 2012. cod. 8992.
Indexations: MEDLINE, Exerpta Medica, Lilacs (Index Medicus Latinoamericano), SciELO (Scientific Electronic Library Online)
CAPES: Qualis Nacional A, Qualis Internacional C


Print:
All Rights reserved © Revista Brasileira de Otorrinolaringologia